Contact us

Tips: * It is a required option. Please make sure that you have filled in all the items to submit
Home > PRODUCTS > Veterinary Health > Diagnostic Examination

Semi-automatic Coagulation Analyzers

Product model:

XNQ6-lV

Detailed introduction:

Model: XNQ6-lVName: Semi-automatic coagulation analyzers (four channels)Features1. Semi-automatic blood coagulation instrument, four channels2. Test items: prothrombin time (PT), activated partial thr... View details >
Message/ Purchase

Product Name: Semi-automatic Coagulation Analyzers (two channels)


Features

1. Semi-automatic blood coagulation instrument, four channels

2. Test items: prothrombin time (PT), activated partial thromboplastin time (APTT), fibrinogen (FIB), thrombin time (TT), other coagulation factors

3. 4 reagent pre-temperature positions and 16 sample pre-temperature positions

4. High brightness backlight LCD display

5. Advanced optical scattering plus percent detection principle, plus automatic zero tracking system technology

6. It can also accurately test special samples containing hyperlipidemia, jaundice, hemolysis and chyle

7. Large-capacity storage, can store 6000 doctor test results, can display and print quality control charts, and make statistics on quality control charts

8. Equipped with electronic guns, with emergency detection function, emergency specimens can be prioritized to prevent external light interference, and it also has power-off protection


Technical Parameters

1. Interface RS-232; built-in thermal printer,

2. Power AC110/220V; 50/60Hz

3. Detection method: optical method

4. ★Test channel: 4

5. ★Specimen type:

(1) Sample pre-temperature position (37°C): 16

(2) Constant temperature pre-temperature position (37°C): 2

6. Temperature control accuracy: ±0.1℃

7. Repeatability error: ≤3% (standard quality control product)

8. Printer: built-in thermal printer

9. Card reader: built-in card reader, which can directly read reagent information

10. Light source: use LED light source

11. Test cup: the cylindrical test cup used

12. Test range: 3 seconds to 300 seconds

13. Sample volume: 50μL~150μL

14. Reagent volume: 50μL~200μL

15. Power supply: 220V±22V, 50Hz±1Hz

16. Working environment temperature: 10℃~35

17. Operating relative temperature: 80%